REMOVE AND REPLACE THE PRODuCT TANK ASSEMBLy (PN 4378-SEQ)

Removal of Product Tank
1) Follow the steps to Removing the Unit Cover

2) Remove the Front Cover Assembly completely; disconnect outlet tubing and front membrane ribbon cable. Set aside.

3) Locate the Product Tank and cut the cable ties from the ATF to the product tank (Bottom), and O2 out from the outlet

connector (Top). Remove tubing from ATF to lower product tank barb. See Figure 36.

Important: Cap off the ATF hose barb with a black hose barb cap.

Figure 36: Cut Cable Ties

4) Locate the 2 screws that hold the securing straps and the product tank in place, remove the screws. Set aside for re-

installation. Figure 37.

Figure 37: Remove Screws

5) Lift out the old product tank. Set aside for RMA return or scrap.

Two screws