Active lifestyle training, Connecting the dc power supply – AirSep Eclipse 5 20631679 User Manual

Page 31

Advertising
background image

Personal Ambulatory Oxygen System Provider Technical Manual

31

PN 20631679 Rev C

When the Eclipse 5 is running from an external (AC or DC) power supply, the External Power Present Indicator will illuminate.
The Eclipse 5 will always operate from external power when external power within the specified range is available.

ACTIVE LIFESTyLE TRAINING

Small, lightweight, and easily moved about, the Eclipse 5 is ideally suited to an active lifestyle. The AC and DC supplies enable
recharge of the Power Cartridge during longer excursions and even extended overnight travel. To train your patient on
mobility and ambulation, follow the procedures below.

Ask your patient whether he or she will be using the device to travel by vehicle. If so, show your patient how to safely use the
Eclipse 5 while driving. Refer to the previous section on Connecting the DC Power Supply for proper setup of the Eclipse 5
system in a vehicle.

CAUTION

When operating an Eclipse 5 5 with the DC power supply, the power
cartridge will only recharge if there is sufficient excess voltage from
what the concentrator requires to operate.

• Ensure adequate clearance around the DC Power Supply and the

Eclipse 5.

• Ensure the DC Power Supply and Eclipse 5 are secured in the vehicle

so that is does not become a projectile in the event of a sudden
stop.

• DO NOT operate the Eclipse 5 on DC power while the vehicle
engine is off. In such a scenario, the Eclipse 5 could quickly drain
the vehicle’s battery.

WARNING:

CONNECTING THE DC POWER SuPPLy

To install the DC Power Supply, follow these steps:

A DC Power Supply allows the system to operate from DC outlets, such as those found in motor vehicles.

1. Consult your patient’s vehicle owner’s manual for DC Accessory outlet requirements.

2. Secure your Eclipse 5 and DC Power supply in your vehicle. Ensure that alerts and alarms are observable.

3. Start the vehicle’s engine.

4. Insert supply plug into DC accessory outlet.

5. Attach the Power Supply plug to the Eclipse 5.

6. Advise the patient to use Pulse Dose Mode if prescribed by their clinician.

7.

When the device is properly connected and receiving power from the DC power source, a green indicator

light on the DC Power Supply will illuminate.

CAUTION

The DC Power Supply is designed for 12VDC minimum vehicle

electrical systems. Do not attempt to operate with 6V, 24V, or other

vehicle electrical system.

Advertising