Eclipse 5 maintenance weekly maintenance—patient, Clean the air inlet filter, Clean and care for the tubing and cannula – AirSep Eclipse 5 20631679 User Manual

Page 35

Advertising
background image

Personal Ambulatory Oxygen System Provider Technical Manual

35

PN 20631679 Rev C

ECLIPSE 5 MAINTENANCE

WEEKLy MAINTENANCE—PATIENT

Training your patient to maintain the Eclipse 5 properly will lead to longer service intervals and lower maintenance costs.
Train your patient to perform the following procedures:

CLEAN THE AIR INLET FILTER

The Air Inlet Filter, located at the rear of the unit, must be cleaned at least once a week.

To clean the filter:

1. Remove the filter from the back of the cabinet.

2. Wash the filter in warm water using a mild detergent

solution.

3. Rinse the filter thoroughly and squeeze out the excess

water.

4. Allow the filter to air dry thoroughly.

5. Re-insert the filter in the cabinet.

6. Maintenance may be easier for the patient when a

second filter is provided.


Figure 10: Cleaning the Air Inlet Filter

CLEAN AND CARE FOR THE TuBING AND CANNuLA

Provide your patient instructions on cleaning, disinfection and/or replacement information for the tubing and cannula.

CLEAN THE CABINET AND CONTROL PANEL AND POWER SuPPLIES

To clean the cabinet, Control Panel or power supply, do the following:

1. Turn OFF the Eclipse 5 and discon nect from AC or DC power before any cleaning or disinfection activity.

2. Use mild detergent and water solution.

3. Use a damp (not soaking wet) cloth or sponge.
4. Spray or wet the cloth or sponge with the mild detergent solution.

DO NOT

spray the cabinet, Control Panel or

power supplies.

5. Wipe down the cabinet, Control Panel or power supplies.
6. To disinfect the Eclipse 5, use Lysol

®

Brand II disinfectant (or equivalent). Spray or wet a cloth or sponge with the

disinfectant.

DO NOT

spray the cabinet, Control Panel or power supplies. Proceed as directed by the manufacturer.

CAUTION

• The filter should be completely dry before using it again. Excess

moisture may impair proper operation or damage the unit.

• Replace the Air Inlet Filter annually. If the Eclipse 5 is operated in a

dusty environment, the filter may need to be cleaned or replaced
more often. Do not operate the Eclipse 5 for more than 30 minutes
without a filter installed.

Advertising