Personal Ambulatory Oxygen System Provider Technical Manual

74

PN 20631679 Rev C

Remove and Replace Power Manager Printed Circuit Board

1 Remove and Replace Combo Power Manager board set

1.1 Remove the Power Cartridge and unscrew the 4 screws (PN: 6974-312-SEQ) holding the BBB pcb to the case shown in Fig 47.

Fig 47: Screws

1.2 Remove the Unit cover as described in section, “Remove and Replace the Unit Cover”.
1.3 Disconnect the 5 wire harnesses as shown in Figure 48.

Fig 48: Cable locations

Disconnect power supplies and remove Power Cartridge before
removing the unit cover. Do not touch exposed circuits during
maintenance without ESD protection.

WARNING:

Compressor Cable

External Power Cable

Cooling Fan Cable

CAUTION

The Eclipse 5 contains electrostatic sensitive components. Do not open
or handle except at a static free workstation.

Compressor Cable