Personal Ambulatory Oxygen System Provider Technical Manual

60

PN 20631679 Rev C

MALFuNCTION CODES

If a malfunction occurs in the Eclipse 5, the device will stop, the Red LED on the front panel will light and the buzzer will sound for 10
seconds and then silence. The LCD will display one of the following Malfunction Codes:

Step 1: Remove all power sources; battery and external power. Wait 20 seconds. Reconnect power to the Eclipse 5. Turn the
device to the ON position. If resolved, continue use.
Step 2: If unresolved, call Chart Technical Support for assistance.

Malfunction

FAIL code

Recommended Action

Invalid RESET

FAIL 80

Check for secure communication cables, Verify functionality of PM and CS PCBs

I/O Port Failure

FAIL 81

Check for secure communication cables, Verify functionality of PM and CS PCBs

RAM Failure

FAIL 82

Check for secure communication cables, Verify functionality of PM and CS PCBs

FLASH Failure

FAIL 83

Check for secure communication cables, Verify functionality of PM and CS PCBs

EEPROM Failure

FAIL 84

Check for secure communication cables, Verify functionality of PM and CS PCBs

IPC Watchdog Timeout

FAIL 90

Check for secure communication cables, Verify functionality of PM and CS PCBs

Compressor Motor Too Hot

FAIL 91

Allow Eclipse 5 to Cool for 30 Mintues. Check compressor fan functionality. Check to ensure vents
are not obstructed. Verify functionality of the Compressor Assy

PCB Too Hot

FAIL 92

Allow Eclipse 5 to cool for 30 Minutes. Verify functionality of PM fan. Check to ensure vents are
not obstructed

Compressor Motor Stalled

FAIL 94

Check Compressor wires for damage. Verify functionality of PM PCB. Verify functionality of the
Compressor Assy

Battery Communication Failure

FAIL 95

Verify functionality of the Battery Bridge PCB. Check Battery and Battery Bridge PCB for pin dam-
age. Verify functionality of battery. Verify functionality of PM pcb

IPC Failure

FAIL A0

Verify functionality of PM PCB

Product Pressure Sensor Failure

FAIL A1

Verify functionality of CS PCB

Breath Pressure Sensor Failure

FAIL A2

Verify functionality of CS PCB

Ultrasonic Failure

FAIL A3

Verify functionality of CS PCB. Check for water intrusion from humidifier

Product Temperature Sensor Failure

FAIL A4

Verify functionality of DB9 cable harness. Verify functionality of CS PCB

Loss of Product Tank Pressure

FAIL A5

Check for leaking, cracked HEPA filter or damaged tubing. Verify ATF motor is turning

Purity Calibration Data Failure

FAIL A6

Check for leaking, cracked HEPA filter or damaged tubing. Verify functionality of CS PCB

Flow Calibration Data Failure

FAIL A7

Verify functionality of Proportional Valve. Verify functionality of CS PCB

Breath Sensitivity Data Failure

FAIL A8

Check for damaged tubing. Verify functionality of CS PCB

Hour of Operation Data Failure

FAIL A9

Verify functionality of CS PCB

Invalid Assembly Option

FAIL AA

Remove Power to Reset

Ambient Pressure Sensor Failure

FAIL AB

Verify functionality of CS PCB

Unidentified Failure

FAIL FF

Remove Power to Reset