Compressor, Remove and replace the compressor box – AirSep Eclipse 5 20631679 User Manual

Page 78

Advertising
background image

Personal Ambulatory Oxygen System Provider Technical Manual

78

PN 20631679 Rev C

NOTE:

There are no field serviceable parts inside the Compressor Box. Do not
attempt to disassemble or modify the Compressor Box in the field.

CAUTION

The Eclipse 5 contains electrostatic sensitive components. Do not open
or handle except at a static free workstation.

Pry

Point

Pry

Point

Pry

Point

Case Bottom

Cover

Compressor

As with any concentrator, the compressor is a limited-wear component and may require servicing
during the lifetime of the device. The point of service will be dependent on factors such as operation
time, flow settings and environmental conditions. Service is required when oxygen purity and/or flow
rates cannot be maintained.

REMOVE AND REPLACE THE COMPRESSOR BOx

1. Remove the Unit Cover as described in the section Remove and Replace the Unit Cover.

2. Flip unit over onto a padded ESD safe surface. Using a small flat tip screwdriver to pry open the cover at each

pry point, remove the Case Bottom Cover as shown in Figure 55.

Figure 55: Case Bottom Cover Removal