Personal Ambulatory Oxygen System Provider Technical Manual

77

PN 20631679 Rev C

1.8 Install the Cooling Fan (PN: 1074-SEQ) with the 2 screws (PN: 6961-210-SEQ) and lightly tighten as shown in Fig 53. Apply Loctite
425 thread locker on the tips of the threads before installing.

Note: Ensure the cooling fan is blowing toward the circuit board as indicated by the arrow on the side of the fan.

Fig 53: Cooling Fan placement

1.9 Turn the device over and align the BBB with the holes in the case. Screw the 4 screws into the Battery Bridge Board from the Battery
compartment side, securing it to the case (snug screws).
1.10 Turn the device back over. Install the 2 clips that come with the power manager board set. Note: There is a right and left clip -
Large gap is on the right - ensure you get both the metal heat sink and pcb corner in the black clip.
1.11 Install the Unit’s front cover according to the “Remove and Replace front cover” section. This holds the Power Manager pcb corner
clips in place (no screw required).

Fig 54: Power Manager installed