Personal Ambulatory Oxygen System Provider Technical Manual

11

PN 20631679 Rev C

Guidance and manufacturer’s declaration–electromagnetic immunity

The Eclipse 5 is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse 5

should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications

equipment should be used no closer to any

part of the Eclipse 5, including cables, than

the recommended separation distance cal-

culated from the equation applicable to the

frequency of the transmitter.

Recommended separation distance

80 MHz to 800 MHz

800 MHz to 2,5 GHz

where P is the maximum output power rating

of the transmitter in watts (W) according to

the transmitter manufacturer and d is the

recommended separation distance in metres

(m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site sur-

vey,

a

should be less than the compliance level

in each frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following sym-

bol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection

from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess

the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the

measured field strength in the location in which the Eclipse 5 is used exceeds the applicable RF compliance level above, the

Eclipse 5 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be

necessary, such as re-orienting or relocating the Eclipse 5.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

d = 1.2 P

d = 1.2 P

d = 2.3 P