ALARM CONDITIONS AND ALARM CODES

Use the table below to decode Eclipse 5 alarm conditions. If other alarm codes are displayed by the Eclipse 5, contact Chart
Technical Support for assistance.

Note: The following table is intended as a guide for the provider, not the user.

CoNDITIoN

ALARM

CoDE

ALARM

DISPLAY

GREEN

INDICAToR

YELLow

INDICAToR

RED

INDICAToR

AUDIBLE

ALARM

wHAT To Do

“Warming Up. Please

Wait.”

004

O2 < 70%

ON

ON

ON

OFF

Wait. The system typically takes 3 – 5 minutes to reach

specified performance.

“Warming Up. Please Wait.”

008

O2 < 85%

ON

ON

OFF

OFF

Wait. The system typically takes 3 – 5 minutes to reach

specified performance.

“3.0” (or flow setting)

000

NO ALARM

ON

OFF

OFF

OFF

Nothing. The system is operating properly at the specified

flow rate.

“Low Power Cartridge”

010

LOW BAT

ON

Flashing

OFF

2 Beeps

Plug into external power or replace with a charged Power

Cartridge.

“Warm Power Cartridge”

002

WARM BAT

ON

Flashing

OFF

1 Beep

Plug into external power or replace with a charged Power

Cartridge.

“Low 9V
Battery”

-

-

OFF

OFF

OFF

3 Beeps on

POST

Replace the internal 9-volt battery.

“O

2

Concentration < 85%”

008

O2 < 85%

OFF

Flashing

OFF

OFF

Perform Annual Filter PM Maintenance. If condition persists,

service is required.

“O

2

Concentration < 70%”

004

O2 < 70%

OFF

OFF

Flashing

3 Beeps

Perform Annual Filter PM Maintenance. If condition persists,

service is required.

“Flow Rate Error / Blocked

Flow”

020

FLOWRATE

OFF

Flashing

OFF

1 Beep

Check tubing and/or humidifier for obstruction. Check/clean

air intake filter. If condition persists past 10 minutes, service

is required.

“Cannot Charge Power

Cartridge”

001

CHARGER

N/A

ON

OFF

OFF

Instruct patient to seek an external power source. Remove
Power Cartridge and allow to cool to room temperature. If

Power Cartridge malfunction persists, service is required.

“One Wire Communica-

tion Loss”

100

ONE WIRE

N/A

Flashing

OFF

1 Beep

Check battery connections.

“No Inspiration Detected”

(in Pulse Mode only)

200

P <--> C

ON

OFF

OFF

OFF

If no inspiratory effort is detected after 15 seconds, the

Eclipse 5 will switch to Continuous Flow for 15 seconds, and

continue this sequence until a breath is detected.

“Loss of External Power”
(without charged Power

Cartridge installed)

040

********
(flashing)

OFF

OFF

Flashing

ON for 5 min

Instruct patient to seek an alternative oxygen supply

until external power is restored or install a charged Power

Cartridge.

“System Fault”

080

FAIL XX

OFF

OFF

ON

ON 10 seconds

then silent

Reset power, wait 20 seconds, reconnect power, restart

device.

Instruct patient to use back up oxygen supply. Service is

required.