Personal Ambulatory Oxygen System Provider Technical Manual

70

PN 20631679 Rev C

Electronics
REMOVE AND REPLACE THE CONTROL BOARD ASSEMBLy

1. Remove the Unit Cover as described in the section Remove and Replace the Unit Cover.

Figure 42: Control Board Removal (ATF Module not shown for clarity).

2. Remove the 2 screws (M3x10 Plastite Screw, P/N 6950-310-SEQ) holding the Control Board Assembly to the

Unit Case as shown in Figure 42.

3. Lift the Control Board Assembly and disconnect the Inlet Air Temperature, Data Communications Port, Power

Manager PC Board Communications, Proportional Valve wire harness, and Buzzer wire harnesses as shown in
Figure 42.

Cable Tie

Screw

Tube from

Product Tank

Tube from

the Proportional

Valve

Buzzer

Connector

Air Inlet

Temperature

Data

Communications

Port

Power

Manager PC Board

Communications

Proportional

Valve Connector

NOTE:

The Control Board Assembly is factory calibrated as a single unit. Do
not disassemble the Control Board Assembly.

CAUTION

The Eclipse 5 contains electrostatic sensitive components.
Do not open or handle except at a static free workstation.

NOTE:

Use a magnetized Phillips #1 screwdriver to remove the screws. This
greatly reduces the chances of dropping a screw in the Unit Case.