Pulse dose mode specifications, Power accessory specifications ac power supply, Dc power supply – AirSep Eclipse 5 20631679 User Manual

Page 7: Power cartridge (battery)

Advertising
background image

Personal Ambulatory Oxygen System Provider Technical Manual

7

PN 20631679 Rev C

Pulse Dose Mode Specifications

Pulse Settings

1.0 to 6.0, in 8mL increments and 128mL, 160mL, 192mL

Trigger Sensitivity

Adjustable between settings of 1 (most sensitive) to 3 (least sensitive)

Adjustable Bolus Rise Time

Adjustable settings of Fast(factory setting), Medium, or Slow

Trigger Criteria

• Cannula pressure has dropped below the trigger point (typically between 0.135—0.37

cm of H

2

O of negative pressure)

• At least 1¼ seconds has passed since the last pulse began

Minimum time between
breaths

1.25 seconds (max. 3 consecutive breaths)

Response to Missing Breaths

While in Pulse Dose Mode, the Eclipse 5 is always monitoring for breath detection.

After 15 seconds of no breath detected, the Eclipse 5 “delivers” Continuous Flow at the

last Continuous Flow setting. After another 15 seconds, the Eclipse 5 stops delivering

Continuous Flow and waits 15 seconds for a breath. The Eclipse 5 will stay in this

modality until a breath is detected. (See Power Cartridge Conservation Feature, page 26.)

Power Accessory Specifications

AC Power Supply

Input Voltage

100-240VAC, 50-60 Hz

Input Power

245-260 VA

Output Voltage

28 VDC

Output Power

200W

DC Power Supply

Input Voltage

11.5-16VDC

Output Voltage

26 VDC

Output Power

150W Max

Power Cartridge (Battery)

Output voltage

14.8 VDC

Capacity

Quantity (2) 97.5 W-hrs batteries
(Each containing 7.92 grams equivalent Lithium content)

Nominal Power Cartridge Life

80% Capacity after 500 Charge/Discharge cycles

Power Cartridge Recharge Time

1.8 to 5.0 hours, dependent on flow setting, to achieve 80% capacity

from a fully discharged Power Cartridge

NOTE: Bolus volume decreases as breath rate exceeds published range.

B

O

Lu

S V

O

Lu

M

E

(+

/-

15

%)

Pulse Dose

Setting

Bolus Size

(± 15%) mL

AC Power Supply and

Power Cartridge (Battery)

DC Power Supply

Max Breath

Rate

Max Breath

Rate

1.0

16

40

40

2.0

32

40

40

3.0

48

40

40

4.0

64

40

31

5.0

80

37

25

6.0

96

31

20

-

128

23

15

-

160

18

12

-

192

15

10

Advertising