Eclipse 5 oxygen system specifications – AirSep Eclipse 5 20631679 User Manual

Page 6

Advertising
background image

Personal Ambulatory Oxygen System Provider Technical Manual

6

PN 20631679 Rev C

Oxygen Concentrator

Dimensions (H x W X D)

19.3 x 12.3 x 7.1 inches

(49.0cm x 31.2cm x 18.0cm)

Weight

Eclipse 5

Power Cartridge

15.0 pounds

3.4 pounds

Flow Settings

Continuous Flow

(measured in Liters Per Minute LPM)

Pulse Dose (measured in mL)

0.5 to 3.0 LPM (0.5 liter increments)

16-96 (8mL increments), 128mL, 160mL, 192mL

Continuous Flow Accuracy

+/- 10% or 200ml/min, whichever is greater

Oxygen Concentration

87–95.6% for all flow settings

Maximum System Pressure

15 psig (103.5 kPa)

Oxygen Output Pressure

5.0 psig (34.5 kPa) nominal

Oxygen Concentration
Status Indicator

Green Light = Normal Operation

Yellow Light = Warning or Caution, less than 85% ± 3%

Red Light Flashing = Abnormal Operation, less than 70% ± 5%

Nominal Sound Level

3.0 LPM Continuous Flow

3.0 Pulse Dose Mode

48 dB(A)

40 dB(A)

Operating Environment

Temperature

Humidity

50º F to 104º F (+10° to 40°C)

10% to 95%, Non-condensing, 82.4°F (28°C) Maximum Dew point

Storage Environment

Temperature

Humidity

-4º F to 140º F (-20° to 60°C)

Up to 95% Non-condensing

Altitude

0 – 13,123 feet (0 - 4,000 meters)

Nominal Power

0.5 LPM Continuous flow

3 LPM Continuous Flow

1.0 Pulse Dose Setting

6.0 Pulse Dose Setting

Battery Charging

48 Watts

145 Watts

45 Watts

95 Watts

50 Watts

Nominal Power Cartridge operating time

Continuous Flow

2.0 LPM

3.0 LPM

Pulse Dose (@ 12 BPM)

2.0

6.0

Continuous Flow Indication

Expressed in liters per minute (LPM)

Audible Alarm Indicators

• Loss of Power/Hot Power Cartridge
• Low Power Cartridge/Warm Power Cartridge
• Low Oxygen Output
• O

2

Flow Outside Normal Limits

• Unit Malfunction
• Low 9V battery

Back-Up Alarm Power

9V Internal Battery

Filters

Air Inlet, HEPA, Compressor Intake

Device Classification

IEC Class I, Type B Applied Part, IPX0

ECLIPSE 5 OxyGEN SySTEM SPECIFICATIONS

Note: Times will decrease with higher bolus size,
breath rate, ambient temperature, power cartridge
age and use over time.

2.0 hours
1.3 hours

5.2 hours
3.5 hours