Selecting Location

Select a place with stable conditions for mounting the transmitter. Do not
expose the transmitter to direct sunlight or rain. A rain shield is available
and it is recommended for direct outdoor installations. When mounting the
probe, select a place representing the process conditions.

General Instructions for Probes with
Cable

Mount the probes with a cable horizontally; this way, any water
condensing on the tube cannot flow onto the sensor.

0507-024

Figure 3

Horizontal Mounting of Probe

When there is no alternative but to install the probe in the process
vertically, the point of entry must be carefully insulated. The cable must
also be allowed to hang loosely as this prevents any condensed water from
running onto the probe along the cable.

The following numbers refer to

Figure 3 on page 20

:

1

=

To be sealed

2

=

To be insulated

3

=

Insulate the cable

4

=

Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable.