Installing the HMP368 Probe Through a Ball Valve
Assembly
The ball valve installation kit (Vaisala order code: BALLVALVE-1) is
preferred when connecting the probe to a pressurized process or pipeline.
Use the ball valve set or a 1/2" ball valve assembly with a ball hole of ø14
mm or more. If you install the probe (ø12 mm) in a process pipe, please
note that the nominal size of the pipe must be at least 1 inch (2.54 cm). Use
the manual press handle to press the probe into the pressurized (< 10 bar)
process or pipeline.
The following numbers refer to
1
=
Probe
2
=
Clasp nut
3
=
Fitting screw
4
=
Pen
NOTE
Take care not to tighten the clasp nut more than 60° to avoid difficulties
when opening it.