Acal analog output calibration – Vaisala HMT360 User Manual

Page 72

Advertising
background image

User’s Guide ______________________________________________________________________

70 ___________________________________________________________________ M010056EN-J

ACAL Analog Output Calibration

Syntax: ACAL<cr>

Connect HMT360 to a multimeter. Issue the ACAL command:

1.

Disconnect the serial cable from the transmitter while reading the
multimeter value for the Ch 1 (I1).

2.

Reconnect the serial cable. Type the multimeter reading and press
Enter.

3.

Type the higher current multimeter reading and press Enter.

Example:

>acal

Ch1 I1 ( mA ) ? 4.846

Ch1 I2 ( mA ) ? 19.987

>

Advertising