User’s Guide ______________________________________________________________________
70 ___________________________________________________________________ M010056EN-J
ACAL Analog Output Calibration
Syntax: ACAL<cr>
Connect HMT360 to a multimeter. Issue the ACAL command:
1.
Disconnect the serial cable from the transmitter while reading the
multimeter value for the Ch 1 (I1).
2.
Reconnect the serial cable. Type the multimeter reading and press
Enter.
3.
Type the higher current multimeter reading and press Enter.
Example:
>acal
Ch1 I1 ( mA ) ? 4.846
Ch1 I2 ( mA ) ? 19.987
>