Classification with current outputs – Vaisala HMT360 User Manual

Page 81

Advertising
background image

Chapter 9 ____________________________________________________________ Technical Data

VAISALA_______________________________________________________________________ 79

Classification with Current Outputs

Europe (VTT)

EU (94/9/EC)

II 1 G Ex ia IIC T4 Ga
VTT 09 ATEX 028 X

Safety factors

U

i

= 28 V, I

i

= 100 mA, P

i

= 700 mW

C

i

= 1 nF, L

i

negligibly low

Environmental specifications

T

amb

= -40 ... +60 °C

Dust classification
(with protection cover 214101)

II 1 D Ex ta IIIC T

500

80 °C Da

-40 °C ≤ T

amb

≤ +60 °C

VTT 04 ATEX 023X

IECEx (VTT)

Ex ia IIC T4 Ga
IECEx VTT 09.0002X

Safety factors

U

i

= 28 V, I

i

= 100 mA, P

i

= 700 mW

C

i

= 1 nF, L

i

negligibly low

Environmental specifications

T

amb

= -40 ... +60 °C

Dust classification
(with protection cover 214101)

Ex ta IIIC T

500

80 °C Da

IECEx VTT 12.0016X

USA (FM)

Classes I, II, III
Division 1, Groups A - G
Division 2, Groups A - D, F and G
FM Project ID: 3010615

Safety factors

V

max

= 28 VDC, I

max

= 100 mA

C

i

= 1 nF, L

i

= 0, P

i

= 0.7 W

T

amb

= 60 °C (140 °F), T5

Japan (TIIS)

Ex ia IIC T4
Code number: TC17897

Safety factors

U

i

= 28 VDC, I

i

= 100 mA, C

i

= 1 nF

P

i

= 0.7 W, L

i

= 0, T

amb

= 60 °C (140 °F)

Advertising