Example:

>asel rh t

Ch1 RH lo : 0.00 %RH

Ch1 RH hi : 100.00 %RH

Ch2 T lo: -40.00 ’C

Ch2 T hi: 100.00 ’C

>

S Scaling Analog Outputs

Syntax: S [zz aa.a bb.b]<cr>

Turn the internal DIP switch Calibration Enabled/Disabled to position
ON before the scaling the analog outputs and return it to position OFF
after scaling.

Example:

>srh 0 100

RH lo: 0.00 %RH

RH hi: 100.00 %RH

>

where
zz

= Quantity (RH, T, Td, x, a, Tw, ppm, aw, RS, Ts)

aa.a = Lower limit of the quantity
bb.b = Upper limit of the quantity