Analog output test, Technical support, Product returns – Vaisala HMT360 User Manual

Page 76: Technical support product returns, Table 7, Error codes, See section, For c, Reports an error; see

Advertising
background image

User’s Guide ______________________________________________________________________

74 ___________________________________________________________________ M010056EN-J

Analog Output Test

Analog outputs can be tested with dip switch 1, see

DIP Switch Functions

on page 45

for details.

Technical Support

For technical questions, contact the Vaisala technical support by e-mail at

[email protected]

. Provide at least the following supporting

information:

-

Name and model of the product in question

-

Serial number of the product

-

Name and location of the installation site

-

Name and contact information of a technically competent person who
can provide further information on the problem.

Product Returns

If the product must be returned for service, see

www.vaisala.com/returns

.

For contact information of Vaisala Service Centers, see

www.vaisala.com/servicecenters

.

Table 7

Error Codes

Error Code
on Display

Error Code
on Serial Line

Typical Cause

RH error

RH error

HUMICAP® humidity sensor open
circut or short circut. Can be
caused by condensed water, or
damage to sensor or probe.

T error

T error

Pt100 temperature sensor open
circut or short circuit. Can be
caused by condensed water, or
damage to sensor or probe.

ADC error

Prb error

Probe module disconnected.

CPU EEPROM error EE error

Hardware failure in transmitter.

PRB EEPROM error EEP error

Hardware failure in probe module.