Water activity in jet fuel applications, Calculated variables (typical ranges), With hmp361 probe – Vaisala HMT360 User Manual

Page 79: Table 10, Table 11, Hmp361 calculated variables specifications, Table 12

Advertising
background image

Chapter 9 ____________________________________________________________ Technical Data

VAISALA_______________________________________________________________________ 77

Water Activity in Jet Fuel Applications

Calculated Variables (Typical Ranges)

With HMP361 Probe

With HMP363, HMP364, HMP365, HMP367 and
HMP368 Probes

Table 10

Water Activity in Jet Fuel Applications Specifications

Property

Description / Value

Measurement range

0 ... 1 a

w

at -40 ... +180 °C (-

40 ... +356 °F)

Accuracy when calibrated against high
quality, certified humidity standards

salt solutions (ASTM E104-85)

± 0.01 (0 ... 0.9)
± 0.02 (0.9 ... 1.0)

± 0.02 (0 ... 0.9)
± 0.03 (0.9 ... 1.0)

Response time (90 %) at +20 °C in still oil
(stainless steel filter)

10 min

Humidity sensor

Vaisala HUMICAP

®

Calculated variables available

mass concentration of water
relative humidity of saturation
saturation temperature

Accuracy of the water content

better than ± 15 % of the
reading

Table 11

HMP361 Calculated Variables Specifications

Property

Description / Value

Dewpoint temperature

-40 ... +60 °C

Mixing ratio

0 ... 160 g/kg d.a.

Absolute humidity

0 ... 160 g/m³

Wet bulb temperature

0 ... +60 °C

Table 12

HMP363, HMP364, HMP365, HMP367& HMP368
Probes Calculated Variables Specifications

Property

Description / Value

Dewpoint temperature

-40 ... +100 °C

Mixing ratio

0 ... 500 g/kg d.a.

Absolute humidity

0 ... 600 g/m³

Wet bulb temperature

0 ... +100 °C