Accuracy of calculated variables, Accuracy of dewpoint temperature °c, Accuracy of mixing ratio g/kg – Vaisala HMT360 User Manual

Page 85

Advertising
background image

Chapter 9 ____________________________________________________________ Technical Data

VAISALA_______________________________________________________________________ 83

Accuracy of Calculated Variables

Accuracy of the calculated variables depend on the calibration accuracy of
the humidity and temperature sensors; here the accuracy are given for ±2
%RH and ±0.2 °C.

Accuracy of Dewpoint Temperature °C

Accuracy of Mixing Ratio g/kg

(Ambient pressure 1013 mbar)

Relative humidity

Temp. 10

20

30

40

50

60

70

80

90

100

-40

1.86

1.03

0.76

0.63

0.55

0.5

0.46

0.43

-20

2.18

1.19

0.88

0.72

0.62

0.56

0.51

0.48

0

2.51

1.37

1

0.81

0.7

0.63

0.57

0.53

0.5

0.48

20

2.87

1.56

1.13

0.92

0.79

0.7

0.64

0.59

0.55

0.53

40

3.24

1.76

1.27

1.03

0.88

0.78

0.71

0.65

0.61

0.58

60

3.6

1.96

1.42

1.14

0.97

0.86

0.78

0.72

0.67

0.64

80

4.01

2.18

1.58

1.27

1.08

0.95

0.86

0.79

0.74

0.7

100

4.42

2.41

1.74

1.4

1.19

1.05

0.95

0.87

0.81

0.76

120

4.86

2.66

1.92

1.54

1.31

1.16

1.04

0.96

0.89

0.84

140

5.31

2.91

2.1

1.69

1.44

1.26

1.14

1.05

0.97

0.91

160

5.8

3.18

2.3

1.85

1.57

1.38

1.24

1.14

1.06

0.99

Relative humidity

Temp. 10

20

30

40

50

60

70

80

90

100

-40

0.003

0.003

0.003

0.003

0.003

0.004

0.004

0.004

-20

0.017

0.018

0.019

0.021

0.022

0.023

0.025

0.026

0

0.08

0.09

0.09

0.1

0.1

0.11

0.11

0.12

0.13

0.13

20

0.31

0.33

0.35

0.37

0.39

0.41

0.43

0.45

0.47

0.49

40

0.97

1.03

1.1

1.17

1.24

1.31

1.38

1.46

1.54

1.62

60

2.68

2.91

3.16

3.43

3.72

4.04

4.38

4.75

5.15

5.58

80

6.73

7.73

8.92

10.34

12.05

14.14

16.71

19.92

24.01

29.29

100

16.26

21.34

28.89

40.75

60.86

98.85

183.66 438.56 —

120

40.83

74.66

172.36 —