Performance

Relative Humidity

Table 8

Relative Humidity Specifications

Property

Description / Value

Measurement range

0 ... 100 %RH

Accuracy (including non-linearity, hysteresis and repeatability)
with HUMICAP® 180
and HUMICAP® 180R
at +15 ... 25 °C

at -20 ... +40 °C
at -40 ... +180 °C

for typical applications
for typical applications
±1 %RH (0 ... 90 %RH)
±1.7 %RH (90 ... 100 %RH)
±(1.0 + 0.008 × reading) %RH
±(1.5 + 0.015 × reading) %RH

with HUMICAP®180L2

at -10 ... +40 °C
at -40 ... +180 °C

for applications with demanding
chemical environment
±(1.0 + 0.01 × reading) %RH
±(1.5 + 0.02 × reading) %RH

Factory calibration uncertainty
(+20 °C)

±0.6 %RH (0 ... 40 %RH)
±1.0 %RH (40 ... 97 %RH)
(Defined as ±2 standard deviation
limits. Small variations possible, see
also calibration certificate.)

Response time (90 %) for
HUMICAP® 180 and HUMICAP®
180L2 at +20 °C in still air

8 seconds with grid filter 20 seconds
with grid + steel netting 40 seconds
with sintered filter

Response time (90 %) for
HUMICAP® 180R at +20 °C
in 0.1 m/s air flow

17 seconds with grid filter
50 seconds with grid + steel netting
60 seconds with sintered filter