Probe mounting, Figure 7, Parts of the transmitter – Vaisala HMT360 User Manual

Page 26

Advertising
background image

User’s Guide ______________________________________________________________________

24 ___________________________________________________________________ M010056EN-J

0603-038

Figure 7

Parts of the Transmitter

Probe Mounting

The following numbers refer to

Figure 7 on page 24

:

1

=

Electronics unit

2

=

Probe; including a part of the measurement electronics (for
example, calibration memory)

3

=

Flat cable

4

=

Transmitter base

5

=

Grounding terminal

6

=

Protective covers

7

=

Cable glands

8

=

Probe

9

=

RS232C connector

CAUTION

Do not unsolder and then resolder the probe cable from and to the printed
board during installation.

Do not shorten or lengthen the probe cable.

These procedures may alter the humidity calibration of the transmitter.