Chapter 4 ________________________________________________________________ Operation

VAISALA_______________________________________________________________________ 47

Selecting Output Quantities

Upper Half of Display

To modify the displayed quantity, turn the Calibration Enabled/Disabled
DIP switch of the transmitter to position Enabled (up). Select the required
quantity with button Up on the display cover, and acknowledge the value
with button E. To complete the selections, turn the DIP switch back to
position disabled (down).

Lower Half of Display

The output quantities metric/non metric DIP switch can be used to select
between °C and °F.

It is possible to check the pressure setting of the transmitter by pressing
button Dn. To return to the temperature reading, press button Dn again.

Character Quantity

Abbreviation

Availability

Metric Unit

Nonmetric Unit

0

relative humidity

RH

A, D

%RH

%RH

1

temperature

T

A, D, F, H

°C

°F

2

dewpoint
temperature

Td

D

°C

°F

3

absolute humidity a

D

g/m

3

gr/ft

3

4

mixing ratio

x

D

g/kg

gr/lb

5

wet bulb
temperature

Tw

D

°C

°F

6

mass
concentration of
water

ppmw

H

ppm

w

ppm

w

7

water activity

aw

H, F

0 ... 1

0 ... 1

8

relative humidity
of saturation

RS

H

%RS

%RS

9

saturation
temperature

Ts

H

°C

°F

The characters used to indicate availability in the table above:
A =

Available for HMT360 standard version

D =

Available for HMT360 with optional calculations

F

=

Available for HMT360 Moisture and Temperature in oil
transmitter

H =

Available for HMT360 Moisture and Temperature in Jet Fuel
transmitter