Chapter 3 _______________________________________________________________ Installation

VAISALA_______________________________________________________________________ 31

The following two fitting body options are available:

-

Fitting Body ISO1/2 solid structure

-

Fitting Body NPT1/2 solid structure

0507-025

Figure 14

Sealing of Fitting Body into Process

Tightening the Clasp Nut

1.

Adjust the probe to a suitable depth according to the type of
installation.

2.

Tighten the clasp nut first manually.

3.

Mark the fitting screw and the clasp nut.

4.

Tighten the nut a further 50 ... 60° (ca. 1/6 turn) with a fork spanner.
If you have a suitable torque spanner, tighten the nut to 45 ± 5 Nm (33
± 4 ft-lbs).

Table 4

HMP368 Probe Dimensions

Probe Type

Probe Length

Adjustment Range

Standard

178 mm

120 mm

Optional

400 mm

340 mm

CAUTION

Take care not to damage the probe body. A damaged body makes the
probe head less tight and may prevent it from going through the clasp nut.