Figure 4, Vertical mounting of probe – Vaisala HMT360 User Manual

Page 23

Advertising
background image

Chapter 3 _______________________________________________________________ Installation

VAISALA_______________________________________________________________________ 21

0507-022

Figure 4

Vertical Mounting of Probe

If the process temperature is much higher than that of the environment, the
whole probe and preferably plenty of cable must be inside the process. This
prevents measuring inaccuracy caused by heat conduction along the cable.

When mounted on the side of a duct or channel, the probe must be inserted
from the side of the duct. If this is not possible and the probe must be
inserted from the top, the point of entry must be carefully insulated.

For Vaisala probe installation kits and some installation examples see
section

Probe Mounting on page 24

.

The following numbers refer to

Figure 4 on page 21

:

1

=

To be sealed

2

=

Insulate the cable

3

=

To be insulated

4

=

Let the cable hang loosely. This prevents condensed water
running to the sensor along the cable