Selecting analog outputs – Vaisala HMT360 User Manual

Page 50

Advertising
background image

User’s Guide ______________________________________________________________________

48 ___________________________________________________________________ M010056EN-J

Selecting Analog Outputs

You can select the output quantities for channels 1 and 2 by turning the
Calibration Enabled/Disabled DIP switch of the transmitter to position
Enabled (up) and the three output selection DIP switches to position
special (all up).

Press button C on the display cover: text "SCAL" appears on the display.
Then press button E until the following display appears:

0505-288

The characters on the second line of the display in this menu correspond to
the quantities according to the table in section

Selecting Output Quantities

on page 47

.

Select the quantity for Ch 1 with buttons Up and Dn and acknowledge the
selection with button E.

If the transmitter is equipped with two analog channels, select the quantity
for Ch 2 in the same way, for example:

0505-289

Press button C to exit the display command mode or continue by setting
the pressure.

NOTE

Remember to restore the DIP switch settings.