Vaisala HMT360 User Manual
Page 39
Chapter 3 _______________________________________________________________ Installation
VAISALA_______________________________________________________________________ 37
Maximum Cable Resistance Calculation for the
Barrier (Vaisala Order Code: 210664)
Calculation of the maximum cable length from barrier to transmitter:
General specifications of HMT360:
Supply voltage
U
in
= 24 V (12 ... 35 V)
Maximum current
I
out
= 20 mA
Minimum operating voltage for
HMT360
U
min
= 12 V (15 V with serial port)
Stahl 9001/51-280-091-141 (values taken from the specifications):
Rated operating voltage
U
N
= 20 ... 35 V
Transmitter supply voltage
U
S
= U
N
- 9.5 V, when
U
N
≤ 23.5 V
or U
S
= 14 V, when U
N
≥ 23.5 V
Maximum load
R
L
≤ 350 Ω
Cable resistance (as an example) R
cable
= 0.085 Ω/m/core (2 ×
0.085 Ω/m/pair)