Troubleshooting questions, Is the ac on led lit, Is the alarm led lit – Apex Digital 5708009-C User Manual

Page 145: Is the overtemp led flashing, Is the interlock led lit, Is the rf on led flashing, Is rf on led lit, Troubleshooting questions -3

Advertising
background image

5708009-C

6-3

Apex 1 to 5.5 kW Generator

Troubleshooting Questions

This section lists the steps necessary to get the unit on. Before working through the
following questions, check the unit’s input and remote power connections to ensure
the proper power is being supplied to the unit. If you have a question regarding
DeviceNet on the Host port call AE Technical Support.

Note: This section uses status LEDs to verify proper operation. If your unit doesn’t

have LEDs, refer to the respective I/O status indicators or the Apex Virtual
Front Panel for verification,

Use the following troubleshooting questions to troubleshoot the Apex generator.

1. Is the AC On LED lit?

– If yes, go to step 2.

– If no, see “AC On LED not Lit” on page 6-4.

2. Is the Alarm LED lit?

– If yes, See “Alarm LED Lit” on page 6-4.

– If no, go to step 3.

3. Is the Overtemp LED flashing?

– If yes, see “Overtemp LED is Flashing” on page 6-4.

– If no, go to step 4.

4. Is the Interlock LED lit?

– If yes, “Interlock LED is not Lit” on page 6-5.

– If no, go to step 5.

5. Is the RF ON LED flashing?

– If yes, see “Alarm LED Lit” on page 6-4.

– If no, go to step 6.

6. Is RF ON LED lit?

– If yes, go to step 7.

– If no, send RF OFF command.