BNC 645 User Manual

Page 104

Advertising
background image

104

-222 Data out of range; large period limits minimum pulse width

The minimum pulse width for pulse waves with period larger than 10
s is larger. Refer to the table of minimum pulse width in Section 4.5
for details.


-222 Data out of range; pulse edge time limited by width; value

clipped to…

The specified pulse edge time does not conform to the following
condition and is adjusted to accommodate the pulse width:

EdgeTime ≤ 0.625 x PulseWidth


-222 Data out of range; pulse edge time limited by duty cycle;

value clipped to…

The specified pulse edge time does not conform to the following
condition and is adjusted to accommodate the duty cycle:

EdgeTime ≤ 0.625 x Period x DutyCycle

-222 Data out of range; period; value clipped to…

The pulse period is out of range and is adjusted.


-222 Data out of range; frequency; value clipped to…

The frequency is out of range and is adjusted.


-222 Data out of range; user frequency; value clipped to upper

limit

The frequency is out of range and is adjusted to the maximum value
allowed due to selection of arbitrary waveform.


-222 Data out of range; ramp frequency; value clipped to upper

limit

The frequency is out of range and is adjusted to the maximum value
allowed due to selection of ramp waveform.


-222 Data out of range; pulse frequency; value clipped to upper

limit

The frequency is out of range and is adjusted to the maximum value

Advertising