8 arbitrary waveform errors, Rbitrary, Aveform – BNC 645 User Manual

Page 110: Rrors

Advertising
background image

110

performing. Use the CAL:SEC:STAT ON command with correct
security code to unsecure the calibration memory.


703 Calibration error; secure code provided was invalid

Security code provided in the CAL:SEC:STAT ON command was
invalid.


706 Calibration error; provide value is out of range

The specified calibration value is out of range.

707 Calibration error; signal input is out of range

The analog-to-digital converter (ADC) verified that the signal applied
to the Modulation In connector on the rear panel is out of range.


707 Calibration error; cal edge time: rise time cal error
707 Calibration error; cal edge time: fall time cal error
707 calibration error; cal edge time: default value loaded

A failure in the rise time or fall time circuitry has prevented calibration
from going forward.

850 calibration error; setup is invalid

Invalid calibration setup number specified.


851 calibration error; setup is out of order

Calibration setup procedures must be performed in a specific
sequence in order to be valid.

7.8 Arbitrary Waveform Errors


770 Nonvolatile arb waveform memory corruption detected

A checksum error was found in the non-volatile memory used to store
arbitrary waveform. As a result, The arbitrary waveform cannot be
retrieved.


781 Not enough memory to store new arb waveform; use

DATA:DELETE

Advertising