10 to enable/disable voltage autoranging, Ontrol, Utput – BNC 645 User Manual

Page 29: Onnector, 11 to control output connector

Advertising
background image

29

Front-panel Operation

1. Press the Utility button and then the softkey under Output Setup.
2. Press the softkey under Rang to toggle between Auto and Hold for

auto-ranging setting.

3. Press the softkey under DONE to finish the selection.

Figure 3-10

Figure 3-11

Remote Interface Operation
VOLTage:RANGe:AUTO {OFF|ON|ONCE}

Notes

• While changing amplitude, you may notice a momentary

disruption in the output waveform at certain voltages due to
switching of the output attenuators. You may use disable the
voltage autoranging to prevent this output disruption.

• With auto-ranging off, the resolution and accuracy of amplitude

and offset may be affected when reducing the amplitude below the
expected range change. And you may not be able to achieve the
minimum amplitude that is available when autoranging is on.

3.11 To Control Output Connector


Model 645 output is controlled by the output button. The output is
disabled at power on by default to protect the equipment connected to
Model 645.

Front-panel Operation

• To enable/disable ouput, press the Output button to toggle. The