6 self-test errors, 7 calibration errors, Alibration – BNC 645 User Manual

Page 109: Rrors

Advertising
background image

109

or associated circuitry.


623 Self-test failed; dc offset DAC
624 Self-test failed; null DAC
625 Self-test failed; amplitude DAC

A possible malfunctioning DAC, failed DAC multiplexer (U803) channels,
or associated circuitry.


626 Self-test failed; waveform filter path select relay
627 Self-test failed; -10 dB attenuator path
628 Self-test failed; -20 dB attenuator path
629 Self-test failed; +20 dB amplifier path

These errors indicate that the specified relay is not properly switched or
the attenuator/amplifier does not provide the expected attenuation or
gain. These self-tests use the internal ADC to verify whether the output
path relays, output amplifier (+20 dB), and output attenuators
properly operate or not.


630 Self-test failed; internal ADC over-range; condition

A probable ADC failure.


631 Self-test failed; internal ADC measurement error

A probable ADC failure.


632 Self-test failed; square/pulse DAC test failure

A probable failure of the square/pulse DAC (U1002).

7.7 Calibration Errors

701 Calibration error; security defeated by hardware jumper

The calibration security feature has been disabled by shorting the two
“CAL ENABLE” pads on the internal circuit board.


702 Calibration error; calibration memory is secured

The calibration memory is secured to disallow a calibration from