-222 Data out of range; duty cycle; value clipped to…

The square wave duty cycle is adjusted to within 20% to 80%.


-222 Data out of range; duty cycle limited by frequency; value

clipped to upper limit

The square wave duty cycle is adjusted to within 40% to 60% when
the frequency is greater than 10 MHz.


-223 Too much data

An arbitrary waveform specified contains more than 65,536 points.
Verify the number of points in the DATA VOLATILE or DATA:DAC
VOLATILE command.


-224 Illegal parameter value

An exact parameter value was expected.

7.3 Device Dependent Errors


-313 Calibration memory lost; memory corruption detected

A checksum error was detected in the non-volatile memory used to
store calibration constant.

-314 Save/recall memory lost; memory corruption detected

A checksum error was detected in the non-volatile memory used to
store instrument states.

-315 Configuration memory lost; memory corruption detected

A checksum error was detected in the non-volatile memory used to
store configuration settings.

-350 Queue overflow

More than 20 errors were found and the error queue was full.