7 to set high-level and low-level bounds, Evel and, Evel – BNC 645 User Manual

Page 26: Ounds

Advertising
background image

26

Figure 3-8

• Limits with amplitude and output termination: DC offset value is

constrained by the equation below:

2

Vpp

Vmax

Voffset

where Vmax is the maximum peak voltage allowed for the
selected output termination (5 volts for a 50 Ω load or 10 volts for
a high-impedance load).
If the specified dc offset is not valid, the waveform generator
automatically adjusts it to the maximum value allowed with the
specified amplitude.

• Arbitrary Waveforms Limitations: For arbitrary waveforms, if the

waveform data points do not span the full range of output DAC
(Digital-to-Analog Converter), the maximum offset and amplitude
will be limited.

• For dc volts, the output level is actually controlled by setting the

offset voltage. You can set the dc level to any value between ±5
Vdc into 50 ohms or ±10 Vdc into an open circuit.


3.7

To Set High-Level and Low-Level Bounds

Setting high-level and low-level bounds provides an equivalent way to
specifying amplitude and DC offset. The procedure is as follows.
Front-panel Operation

1. Select your desired function by pressing one of the function keys.
2. Press the softkey under Ampl/High to toggle and make sure the

High is highlighted. Use the numeric keypad or the knob with
cursor keys to change the value.

3. Press the softkey under Vos/Low to toggle and make sure Low is

highlighted. Use the numeric keypad or the knob with cursor keys
to change the value.

Advertising