7 calculation, 1 general, Calculation – Metrohm 877 Titrino plus User Manual

Page 100

Advertising
background image

7.2 Endpoint titrations (SET)

■■■■■■■■■■■■■■■■■■■■■■

90

■■■■■■■■

877 Titrino plus

Input range

0.00000 - 9999.99 mL

Default value

100.000 mL

Selection

off

Stop time

The titration is stopped when the specified time has elapsed following the
termination of the start conditions.

Input range

0 - 999999 s

Selection

off

Default value

off

Filling rate

Rate at which the dosing cylinder is filled after the titration. The maximum
filling rate depends on the cylinder volume (see Chapter 10.1.1, page
101)
.

Input range

0.01 - 166.00 mL/min

Selection

max.

Default value

max.

7.2.7

Calculation

7.2.7.1

General

Menu

Parameters Calculation

A maximum of five calculations can be defined in one method. A series of
variables (raw data from the determination, previously calculated results) is
available for the calculations. Definition can be made for each calculation
as to whether the result is to be saved as a titer or as a common variable.

The result name is specified in the list for each calculation.

Edit

Edit the data of the selected calculation, see following chapter.

Delete

Delete the selected calculation.

Advertising