6.6 Configuring external devices
■■■■■■■■■■■■■■■■■■■■■■
64
■■■■■■■■
877 Titrino plus
The interface parameters for the connected balance are set under COM1
settings.
Baud rate
Transfer rate in characters per second.
Selection
1200 | 2400 | 4800 | 9600 | 19200 | 38400 |
57600 | 115200
Default value
9600
Data bits
Number of data bits.
Selection
7 | 8
Default value
8
Stop bits
Number of stop bits.
Selection
1 | 2
Default value
1
Parity
Type of parity testing.
Selection
even | none | odd
Default value
none
Handshake
Type of the data transfer protocol.
Selection
hardware | software | none
Default value
hardware
NOTE
If communications problems occur, set the parameter Handshake to
software, and make another attempt.
Editing the COM2 settings
Menu
▶ System ▶ External devices ▶ COM2 settings
The interface parameters for instruments connected to the RS-232/2 con-
nector of the RS-232/USB Box (e.g. PC) are set under COM2 settings.
The parameters and input ranges are identical to those for the COM1
interface.