5.13 Manual Control
■■■■■■■■■■■■■■■■■■■■■■
44
■■■■■■■■
877 Titrino plus
NOTE
The Stirrer function can only be selected if a stirrer is connected.
5.13.1
Dosing
The following dosing functions are available in the manual control:
Prepare buret unit
(PREP)
Rinse the cylinder and tubings of the buret unit
(see Chapter 10.1.2, page 101).
Continuous dosing
(DOS)
Dose while the [START] key is pressed.
Dose fixed volume
(ADD)
Dose a specified volume.
Preparing the buret unit (PREP)
The PREP function is used to rinse the cylinder and tubings of the buret
unit and fill them air bubble-free. You should carry out this function
before the first determination or once per day.
Proceed as follows:
1 Open the manual control
■
In the main dialog, select Menu and press [OK].
The main menu opens.
■
Select the menu item Manual control and press [OK].
Manual control opens.
2 Select the dosing function
■
Select the Dosing item.
■
In the function bar, select PREP and press [OK].
The following message is displayed: