Metrohm 877 Titrino plus User Manual

Page 93

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Parameters

877 Titrino plus

■■■■■■■■

83

Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Dosing rate

Rate at which the start volume is dosed. The maximum dosing rate
depends on the cylinder volume (see Chapter 10.1.1, page 101).

Input range

0.01 - 166.00 mL/min

Selection

max.

Default value

max.

Pause

Waiting time, e.g. in case of a slow response of the electrode after the
start or reaction time after the dosing of a start volume.

Input range

0 - 999999 s

Default value

0 s

Request sample ID

Selection of the sample identification that is queried in the method run.

Selection

off | ID1 | ID2 | ID1&ID2

Default value

off

Request sample size

If this parameter is switched on, then the value for the sample size will be
queried.

Selection

on | off

Default value

off

Request sample unit

If this parameter is switched on, then the unit for the sample size will be
queried.

Selection

on | off

Default value

off

Hold at request

If this parameter is switched on, then the method run will be paused dur-
ing the query. If the parameter is switched off, the titration will be started
in the background.

Selection

on | off

Default value

on

Advertising