7.2 Endpoint titrations (SET)

■■■■■■■■■■■■■■■■■■■■■■

84

■■■■■■■■

877 Titrino plus

7.2.3

Titration parameters

Menu

Parameters Titration parameters

The parameters that influence the sequence of the entire titration are
defined under Titration parameters.

Solution

Selection of the solution from the solution list. We recommend always
selecting the solution. This ensures that the correct data (titer, concentra-
tion, etc.) is always used for the calculation. Solutions are defined under
System

Solutions.

For buret units with integrated data chip, a check is made in the method
run to verify whether the correct solution has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. The
validity of the titer is checked for the selected solution at the start of the
determination.

Selection

Selection of configured solution | not defined

Default value

not defined

not defined
No check takes place.

Sensor

Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System

Sensors.

Selection

Selection of configured sensors

I(pol)

The polarization current is the current that is applied at a polarizable elec-
trode during the voltametric measurement. This parameter is available
only with I(pol) determinations.

Input range

–125 - 125 µA (Increment: 1)

Default value

5 µA

U(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement. This parameter is available only
with U(pol) determinations.

Input range

–1250 - 1250 mV (Increment: 10)

Default value

400 mV