Metrohm 877 Titrino plus User Manual

Page 66

Advertising
background image

6.2 Managing sensors

■■■■■■■■■■■■■■■■■■■■■■

56

■■■■■■■■

877 Titrino plus

Slope

This parameter only appears for pH electrodes.

Slope of the pH electrode. With a 1-point calibration, only pH(0) can be
calculated, 100.0 % is used as the slope.

Input range

–999.9 - 999.9 %

Default value

100.0 %

pH(0)

This parameter only appears for pH electrodes.

pH value of the pH electrode at 0 mV. Apart from the slope, pH (0) is the
second characteristic of the calibration curve.

Input range

–20.000 - 20.000

Default value

7.000

Calibration temp.

This parameter only appears for pH electrodes.

Temperature at which the last calibration was carried out.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Calibration date

This parameter only appears for pH electrodes.

Date of the last calibration.

Monitoring

This parameter only appears for pH electrodes.

Activating and deactivating the calibration monitoring.

Selection

on | off

Default value

off

Time interval

This parameter is visible only when Monitoring = on.

You will be notified that this time interval (in days) has elapsed when start-
ing a method. You can then select whether or not you would still like to
start the method.

Input range

1 - 999 d

Default value

999 d

Advertising