Metrohm 877 Titrino plus User Manual

Page 98

Advertising
background image

7.2 Endpoint titrations (SET)

■■■■■■■■■■■■■■■■■■■■■■

88

■■■■■■■■

877 Titrino plus

Measuring mode Upol:
Input range

0.01 - 125.00 µA

Default value

10.00 µA

Selection

off

Max. rate

This parameter is visible only when Titration rate = user.

Rate at which dosing is carried out outside of the control range. The maxi-
mum dosing rate depends on the cylinder volume (see Chapter 10.1.1,
page 101)
.

Input range

0.01 - 166.00 mL/min

Default value

10.00 mL/min

Selection

max.

Min. rate

This parameter is visible only when Titration rate = user.

Rate at which dosing is carried out at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.

Input range

0.01 - 9999.00 µL/min

Default value

25.00 µL/min

Table 2

Standard values for predefined titration rates for SET

Titration rate

slow

optimal

fast

Dynamics

– pH

– U und Ipol

– Upol

5.000

300.0 mV

40.00 µA

2.000

100.0 mV

10.00 µA

0.500

30.0 mV

5.00 µA

Max. rate

1.00 mL/min

10.00 mL/min

maximum

Min. rate

5.00 µL/min

25.00 µL/min

50.00 µL/min

Stop criterion

The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see Chapter 7.2.6,
page 89)
always lead to a stop, even if the stop criterion has not been
reached.