2 dosing drive, 3 interfaces, 4 mains connection – Metrohm 877 Titrino plus User Manual

Page 126: 5 safety specifications

Advertising
background image

11.2 Dosing drive

■■■■■■■■■■■■■■■■■■■■■■

116

■■■■■■■■

877 Titrino plus

11.2

Dosing drive

Resolution

10,000 steps per cylinder volume

Exchange unit

Cylinder volume

1 mL

5 mL

10 mL

20 mL

50 mL

Accuracy

Fulfills ISO/DIN standard 8655-3

11.3

Interfaces

USB (OTG) connec-
tor

For connecting USB devices.

MSB connector

For connecting a stirrer.

Remote connector

For connecting instruments with a remote interface.

11.4

Mains connection

Voltage

100 - 240 V (± 10%)

Frequency

50 - 60 Hz

Power consump-
tion

45 W

Fuse

2 × 2.0 ATH, electronic overload protection

11.5

Safety specifications

This instrument fulfills the following electrical safety requirements:

CE designation in accordance with the EU directives:

2006/95/EC (Low Voltage Directive, LVD)

2004/108/EC (EMC Directive, EMC)

Federal Inspectorate for Heavy Current Installations ESTI (Accreditation
Number SCESp 033)

Safety mark for certification type 2 in accordance with NEV (type
testing with market monitoring, EMC conformity)

Advertising