Metrohm 877 Titrino plus User Manual
Page 117
Advertising
■■■■■■■■■■■■■■■■■■■■■■
10 Appendix
877 Titrino plus
■■■■■■■■
107
Table 6
Inputs and outputs of the remote interface
Pin No.
Assignment
Function
1
Output 0
Ready/EOD
2
Output 1
Activate/Dosimat
3
Output 2
Titration
4
Output 3
Cond OK
5
Output 4
Error
6
0 volt (GND)
7
+5 volts
8
Input 0
Start
9
Input 1
Stop
Inputs
+5 V
approx. 5 k
Ω Pull-up
tp
t
p
> 100 ms
active = low, inactive = high
Outputs
Open Collector
tp
t
p
> 200 ms
active = low, inactive = high
I
C
= 20 mA, V
CEO
= 40 V
+5 V: maximum load = 20 mA