Metrohm 761 Compact IC User Manual

Page 140

Advertising
background image

4 Operation

761 Compact IC

130

Sample

Sample information from the

Sample

tab of the

passport.

SAMPLE: Tap water from Application Laboratory
: Herisau, Switzerland
Vial number: 2
Volume: 20.0 µl
Dilution: 1.00

Column

Column information from the

Column

tab of the

passport.

COLUMN: METROSEP Anion Dual 2 (6.1006.100)
Size: 4.6 x 75 mm
Number: A106
Part.size: 6.0 µm

Eluent

Eluent information from the

Eluent

tab of the

passport.

ELUENT: 2.0 mM NaHCO3 / 1.3 mM Na2CO3


Flow: 0.80 mL/min
Temperature: 20.0°C
Pressure: 3.4 MPa

Chromatogram plot

Plot of the chromatogram (only for

Printer

or

File

destination).

Peak table

Peak table with results. The structure of the

peak table depends on the selected

Quantifica-

tion method

. The following report is put out as

default:

Quantification method: Custom
Standard component: No
Normalization: 100.00

No Retention Height Area Conc. Name
min uS/cm uS/cm*sec mg/L
1 2.83 0.09 0.803 0.04 fluoride
2 5.06 5.35 94.134 8.06 chloride
3 6.11 0.31 10.254 0.00
4 10.57 2.52 57.007 9.06 nitrate
5 16.87 1.43 53.565 6.40 sulfate
_____________________________________________

5 20.01 9.70 215.763 23.57