Metrohm 761 Compact IC User Manual

Page 34

Advertising
background image

2 Installation

761 Compact IC

24

Select the

Connect to workplace

menu item of the

Control

menu

in this window.

If the connection between PC and Compact IC is working the

message

Hardware settings file for 761 unit with serial number

'#####' not found! Create?

will appear. Click

<Yes>

in order to

create the configuration file

#####.761

for this instrument.

If the connection between PC and Compact IC does not work

then the message

Detection of hardware failed[761 Compact IC

[ COM# ]]

appears in the

SYSTEM STATE

window. In this case

check whether the instrument has been switched on, whether

the connection cable is connected up properly and whether

the COM interface has been set correctly (see point 2). Then

repeat point 3.

4 Hardware settings

Of the general hardware settings only the input of the cell constant

is described. Standard settings can normally be used for all other

parameters.

Click the 761 icon using the right mouse button and select

the

Hardware

item. The

Hardware settings

window is opened:

In the

Cell constant

field enter the cell constant which is printed

on the 1.732.0110 Detector block (see section 2.3.1).

Click on

<OK>

to close the window and save the settings.

Advertising