2.7 Precolumns and separating columns

761 Compact IC

35

2.7.5

General information on separating columns

New IC separating columns are normally filled with solution and

sealed at both ends. Before the column is installed in the system, it

must be ensured that this solution is freely miscible with the eluent

used (check manufacturer's specifications).

The IC separating columns and precolumns currently available from

Metrohm are listed in section 6.3.2. A test chromatogram and an infor-

mation leaflet is provided with each column. You will find additional in-

formation concerning these columns in the 8.732.2003 Metrohm Mono-

graph «Ion chromatography» and in special "Application Bulletins",

which are available on request free of charge from your local Metrohm

agency.

When you install the column, always ensure that this is inserted

correctly in accordance with the flow direction shown (arrow must

point upwards).






2.7.6

Selection of the sample loop

Selection of the sample loop depends on the separating column used.

Normally, the following sample loops are used:

Cation columns

10

µ

L

Anion columns with suppressor

20

µ

L

Anion columns without suppressor

10

µ

L

Depending on the instrument version, the following sample loops are

installed in the 761 Compact IC 761:

Version

Sample loop

Volume

2.761.0010

6.1825.230 (PEEK)

10

µ

L

2.761.0020

6.1825.210 (PEEK)

20

µ

L

If desired, the built-in sample loop can be replaced by one of the sam-

ple loops available as an option (see section 6.3.1).