Metrohm 761 Compact IC User Manual

Page 194

Advertising
background image

5 Notes – Maintenance – Faults

761 Compact IC

184

3 Clean suppressor rotor

Clean the sealing surface of new suppressor rotor 127

(6.2832.000) using a lint-free cloth and ethanol.

4 Insert suppressor rotor

Insert new suppressor rotor 127 in suppressor holder 128 in

such a way that the tubing connections at the rear of the rotor

fit in the corresponding openings inside the rotor and that one

of the three holes in the rotor can be seen from below in one

of the openings of the holder.

If the rotor has been inserted correctly, its sealing surface will

be about 4 mm inside the holder. If this is not the case, bring

the rotor into the correct position from below with the aid of a

sharp object (e.g. a screwdriver).

5 Clean connection piece

Clean the sealing surface of new connection piece 126

(6.2832.010) with the aid of a lint-free cloth and ethanol.

6 Insert connection piece

Insert new connection piece 126 in suppressor holder 128 in

such a way that connection "1" is at the top and that the three

lugs on the connection piece fit in the corresponding open-

ings of the holder.

Screw nut 125 onto the thread of suppressor holder 128

manually (do not use tools).

7 Connect and condition the suppressor

Reconnect the suppressor to the IC system.

Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.

Advertising