4 Operation

761 Compact IC

158

Started

Indication whether the sample has been started

or not:

0

sample not started

1

sample started

2

n

number of injections done

Level

Calibration level (see section 4.4.5) for the

sample: Level

0

stays for normal analysis, levels

1

and greater for calibration runs. Correct filling

of this column enables to calculate calibration

coefficients automatically.

Volume

Injected volume of sample in

µ

L.

Amount

Sample amount. If this value is different for the

calibration run (c) and the sample run (s), the

component concentrations of the sample are

calculated as follows:

C

s

=

C

c

Amount

s

/

Amount

c

Shut down system after the queue finishes

If this option is enabled, the system hardware is

shut down after the sample queue has been fin-

ished (all pumps are switched off), but it re-

mains connected.

<Start>

Start execution of the sample queue from the

first line with

Started = 0

.

<Pause>

Stop execution of the sample queue after the

current determination has been finished.

<Edit>

Open the sample queue editor program for ed-

iting the sample queue table (see section

4.6.3).

<Close>

Close the sample queue overview table.



Start sample queue

SAMPLE QUEUE OVERVIEW /

Control / Start

Start execution of the sample queue from the first row with

Started = 0

.

For each row, the sample-specific data are transferred to the method of

the current determination and the parameter

Started

is set to

1

for this

row.

If processing the sample queue is interrupted with

<Pause>

then it can

be continued again with

<Start>

(with a running chromatogram only

when data acquisition is finished).