6.1 Technical data

761 Compact IC

189

6 Appendix

6.1 Technical

data

6.1.1 Conductivity measurement

Measurement range 1

0

1000

µ

S/cm (resolution: 0.56 nS/cm)

Measurement range 2

0

250

µ

S/cm (resolution: 0.14 nS/cm)

Measurement range 3

0

50

µ

S/cm (resolution: 0.028 nS/cm)

Maximum error

±

1 % of full scale value and

±

1 % of measurement value (k = 16.7/cm)

Linearity

Deviations <

±

0.5 % of full scale value

Noise

Measurement range 1

Measurement range 2

Measurement range 3

typ. 10 nS/cm

typ. 2.5 nS/cm

typ. 0.5 nS/cm

Drift (electronic)

typ. < 10 ppm/h of full scale value

Temperature dependence

typ. < 40 ppm/°C of full scale value

Reserve range

> 33 % (k = 16.7/cm)

Sampling rate

10 measurements/s (fixed)

6.1.2 Conductivity detector

Construction

Thermostatted conductivity detector with

2 ring-shaped steel electrodes

Measurement principle

Alternating current measurement with 1 kHz

frequency and ca. 1.7 V amplitude (peak to

peak).

Effective cell volume

0.8

µ

L

Cell constant

approx. 17 /cm (the exact value is printed on the

detector)

Maximum back pressure

for measuring cell

5.0 MPa (50 bar)

Thermostatting

Connectable dynamic control to adjustable

operating temperature

Operating temperature

Adjustable in steps of 5°C from 25

45°C

Max. temperature deviation

±

2.5°C

Heating time

30 min

Temperature stability

0.01°C at constant ambient temperature

Connection for detector block Dsub 15 pin (female)