5.3 Faults and malfunctions

761 Compact IC

187

Malfunction

Cause

Rectification

Chromatograms
with poor resolu-
tion, change in the
retention times

Deterioration in
separation efficiency of
the IC column

Regenerate column (see
section 5.1.1) or replace
column

Extreme peak
broadening,
splitting (double
peaks)

Dead volume at the
column ends

Fill column with glass
beads (

100

µ

m) or

replace column

No feed of
regeneration or
rinsing solution
for suppressor

Contact pressure too
low

Leak in system

Defective pump tubing

Contamination of the
filter in the 6.2821.120
Filter unit PEEK

Counterpressure at
suppressor module too
high

Adjust contact pressure
(see section 2.9.2)

Check connections and
make leakproof

Replace pump tubing (see
section 5.2.9)

Replace the 6.2821.130
Filter (see section 2.3.6)

Clean or replace suppres-
sor (see section
5.2.6

5.2.8)